Wave Quantities

The Wave Quantities submenu selects a wave quantity to be measured and displayed.

Wave quantities

In contrast to S-parameters, wave quantities are not system error corrected. A power calibration can be applied to wave quantities; see Data Flow.  


a1 Src Port 1, b1 Src Port 1, b2 Src Port 1, a2 Src Port 2, b1 Src Port 2, b2 Src Port 2

Select the standard 2-port wave quantities a1, a2, b1, and b2 for different source ports:

The predefined wave quantities are obtained with different source ports. a1 Src Port 1, b1 Src Port 1 and b1 Src Port 2 are measured at Port 1 of the analyzer. a2 Src Port 2, b2 Src Port 1 and b2 Src Port 2 are measured at Port 2 of the analyzer.  

The analyzer can also measure wave quantities for other source ports; see More Wave Quantities.

Remote control:

CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "A1" | ...  
[SENSe<Chn>:]FUNCtion[:ON] "...
POWer:A1" etc.

Create new trace and select name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "A1" | ...
 


More Wave Quantities

Opens a dialog to define and select arbitrary wave quantities for different detectors and source ports or more than 2 ports.   

The notation for wave quantities and the functionality of the More Wave Quantities dialog is analogous to the definition of S-parameters.

The measurement process for external generators Gen 1, Gen2... in the Source Port list differs fromthe measurement process for internal source ports:

  1. An external generator always represents a permanent signal source that is switched on for all partial measurements. In contrast, an analyzer port is switched off for partial measurements that do not require a source signal. Exception: Internal ports can be configured as permanent sources (see Gen setting in theSourcesection of the Port Configuration dialog).

  2. The external source is measured in the first partial measurement where an internal source is active. This means that no separate partial measurement for the external generator signal is needed. If no internal source is needed at all, the external source is measured in the first partial measurement.  

Remote control:

CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "A1" | ...  
[SENSe<Chn>:]FUNCtion[:ON] "...
POWer:A1" etc.

Create new trace and select name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "A1" | ...